Waltron 3045 Copper Analyzer

DISCONTINUED: The 3045 Dissolved Copper Analyzer is replaced by the 3145 Dissolved Copper Analyzer


3045 Dissolved Copper Analyzer

Range: 0–100ppb

Range: 0–3ppm


The Waltron 3045 Dissolved Copper Analyzer is the most reliable instrument for online colorimetric measurement of dissolved copper in clean water applications. Online dissolved copper analysis is increasingly critical as an effective method to monitor corrosion within specialty steam and boiler water systems. An enhanced color touchscreen user interface makes the 3045 easy to operate and configure, and the clever fast-loop reservoir feature allows accurate measurement independent of flow. The 3045 alone can deliver the most accurate and precise results for ultra-pure water applications.

  • Accurate and wide-range analysis

  • Multi-stream available

  • Fast loop reservoir

  • Adjustable cycle and calibration frequency

  • Automatic calibration and validation (QA/QC)

  • Compact design

  • Color touchscreen display

  • Grab sample capability

  • Internal datalogger

  • Low reagent level alarm

  • No compressed air is required

  • Automatic start/stop based on sample flow detection

  • Minimal analyzer maintenance

PERFORMANCE


Range

0–100 ppb, 0–3 ppm


Accuracy

+/– 2% of full scale


Reproducibility

+/– 2% of reading


Analysis Time

10 min, batch process


Sampling Frequency (Cycle Time)

One reading every 15 mins, adjustable


Sensor Classification

Colorimetric with glass flowcell and LED source


Power Supply

110-220Vac, 50-60 Hz 80 VA


Humidity

Up to 90% not condensable


Ambient Temperature Range

5-55°C analyzer (41-131°F)


Dimensions

Height = 26” (66.1cm), Width = 15” (38.1cm) Depth = 12” (30.5 cm)


Positioning and Installation Details

Wall mounting, bench supported, or panel mounting (optional)


Alarms

2 or 4 configurable alarm relays


Degree of Protection

IP55 (NEMA 4): Wet section, IP65 (NEMA 4x): Electronics


Required Maintenance

Once a month replenish reagents, Quarterly replace tubing


Reagent Consumption (approximate)

3 reagents, 2L each per month


Materials in Contact with Sample

Glass, Silicone, Plexiglass, Stainless Steel AISI 316


Certification (on request):

Meets low voltage and low electromagnetic compatibility directives


Sample Conditioning Requirements

Filtering between 10 and 60 micron, depending on the matrix, needed only to avoid clogging.


Weight

17 kg (37.5 lbs.)

SAMPLE DELIVERY OPERATING CONDITIONS


Temperature Range

5-55º C (41-131º F)


Inlet Sample

100–500 ml/min, atmospheric, flexible tubing O.D. 1/4″ (6 mm)


Outlet Sample

Atmospheric, waste tubing O.D. 3/8 (9 mm)


PH

3-12

SIGNAL OUTPUTS


Analog Outputs

4-20mA (galvanic isolator module available as option) or 0-5V


Serial I/O for Signals

RS232, RS485; MODBUS; PROFIBUS

OPERATIONAL CALIBRATION


Frequency/Intervals

Recommended: 7 days


Single/Multi-Point

Multi-Point: zero and mid-range


Manual/Automatic

Both


TECHNICAL DOCUMENTS