DISCONTINUED: The 3048 Iron Analyzer is replaced by the updated 3148 Dissolved Iron Analyzer.
3048 Iron Analyzer
Range: 0 – 250 ppb
Range: 0 – 500 ppb
Range: 0-5 ppm
The Waltron 3048 Dissolved Iron Analyzer is the first choice for colorimetric online analysis of dissolved iron in clean water applications. Online dissolved iron analysis is increasingly critical as an effective method to monitor corrosion within steam and boiler water systems. An enhanced color touchscreen user interface makes the 3048 easy to operate and configure, and the clever fast-loop reservoir feature allows accurate measurement independent of flow. The 3048 automates manual bench test analysis for accurate and reliable results, greatly enhancing visibility on iron concentration changes and earlier detection of potential problems.
- Accurate and wide-range analysis
- Multi-stream available
- Fast loop reservoir
- Adjustable cycle and calibration frequency
- Automatic calibration and validation (QA/QC)
- Compact design
- Color touchscreen display
- Grab sample capability
- Internal datalogger
- Low reagent level alarm
- No compressed air is required
- Automatic start/stop based on sample flow detection
- Minimal analyzer maintenance
PERFORMANCE
Range
0-250 ppb, 0-500 ppb, 0-5 ppm (internal dilution)
Accuracy
+/– 2% of full scale
Reproducibility
+/– 2% of reading
Analysis Time
8 min, batch process
Sampling Frequency (Cycle Time)
One reading every 20 minutes (adjustable)
Required Maintenance
Once a month, replenish reagents, replace tubing quarterly
Reagent Consumption (approximate)
1 reagent, 1L per month
Sensor Classification
Colorimetric with glass flowcell and LED source
Application
Boiler water and cooling water
Ambient Temperature Range
5-55°C analyzer (41-131°F)
Atmospheric Pressure Range
No Limits
Humidity
Up to 90% not condensable
Unit Dimensions
Height=23.6”(60 cm), Width=15”(38 cm), Depth=8.3”(21 cm)
Positioning and installation details
Wall mounting or with bench support, panel mounting (optional)
Weight
17 kg (37.5 lbs.)
Degree of Protection
IP55 (NEMA 4): Wet section, IP65 (NEMA 4x): Electronics
Alarms
2 or 4 configurable alarm relays
Power Supply
110-220Vac, 50-60 Hz 80 VA
CE Certification (on request)
Meets low voltage and low electromagnetic compatibility directives
SAMPLE DELIVERY OPERATING CONDITIONS
Temperature Range
5-55º C (41-131º F)
Inlet Sample
100–500 ml/min, atmospheric, flexible tubing O.D. 1/4”(6 mm)
Outlet Sample
Atmospheric, waste tubing O.D. 3/8”(9 mm)
PH
3-12
Sample Conditioning Requirements
Filtering between 10 and 60 micron, depending on the matrix, needed only
to avoid clogging
Materials in Contact with Sample
Glass, Silicone, Plexiglass, Stainless Steel AISI 316
SIGNAL OUTPUTS
Analog Outputs
4-20mA (galvanic isolator module available as option) or 0-5V
Serial I/O Options
OPERATIONAL CALIBRATION
Single/Multi-Point
Multi-Point: zero and mid-range
Manual/Automatic
Both