3052 Disinfectants Analyzer
Range: 0-5 ppm
Waltron’s 3052 Disinfectants Analyzer is the first choice for online analysis of monochloramine. The 3052 is trusted worldwide to deliver accurate and reliable readings with minimal maintenance and a low total cost of ownership. Typical applications are potable water, cooling water, industrial wastewater, and swimming pools.
- Accurate and wide-range analysis
- Multi-stream available
- Fast-loop reservoir
- Adjustable cycle and calibration frequency
- Automatic calibration and validation (QA/QC)
- Compact design
- Color touchscreen display
- Grab sample capability
- Internal datalogger
- Low reagent level alarm
- No compression air required
- Automatic start/stop based on sample flow detection
- Minimal analyzer maintenance
- Process-photometer based on DPD-Method (DIN EN ISO7393-2; APHA/AWWA 4500-CI G)
Range
0-5 ppm
Accuracy
± 5% of the reading
Reproducibility
± 0.1 ppm or ± 2% of reading, whichever is greater
Analysis Time
10 min batch process
Sampling Frequency (Cycle Time)
One reading every 20 mins, adjustable
Required Maintenance
Monthly replenish reagents; Quarterly replace tubing
Reagent Consumption (approximate)
2 reagents, 2.5 L each month
Sensor Classification
Colorimetric with glass flow cell and LED source
Application
Steam cycle water, cooling water, process water
Ambient Temperature Range
10-45°C analyzer (50-113°F)
Atmospheric Temperature Range
No limits
Humidity
Up to 90% not condensable
Unit Dimensions
Height = 23.6″ (60 cm); Width = 15″ (38 cm); Depth = 8.3″ (21 cm)
Positioning and Installation Details
Wall mounting, bench supported, or panel mounting (optional)
Weight
17 kg (37.5 lbs)
Degree of Protection
IP55 (NEMA 4): Wet section, IP65 (NEMA 4x): Electronics
Alarms
2 or 4 configurable alarm relays (result, loss of sample, fault)
Power Supply
110-220 VAC, 50-60 Hz 80 VA
Certification (on request)
Meets low voltage and low electromagnetic compatibility directives
Temperature Range
5-55°C (41-131°F)
Inlet Sample
100-500 ml/min, atmospheric, flexible tubing O.D. 1/4″ (6mm)
Outlet Sample
Atmospheric, waste tubing O.D. 3/8″ (9mm)
pH
3-12
Sample Conditioning Requirements
Filtering between 10 and 60 micron, depending on the matrix, needed only to avoid clogging
Materials in Contact with Sample
Glass, Silicone, Plexiglass, Stainless Steel AISI 316
Analog Outputs
4-20 mA (galvanic isolator module available as option) or 0-5V
Serial I/O Options
RS232; RS485; MODBUS; PROFIBUS
Single/Multi-Point
Multi-point: zero and mid range
Matrix Corrections
Yes, sample blank correction
Manual/Automatic
Both