4000 SC Specific Conductivity Analyzer

Ranges from 0-9.999µS/cm to 0-999.9mS/cm


Streamline Water and Steam Monitoring with Waltron’s All-New Conductivity Analyzers

Waltron’s 4000 SC Specific Conductivity Analyzer is designed for definitive water and steam quality monitoring. This high-precision transmitter boasts a broad specific conductivity range from ultra-pure to general-purpose water, ensuring its versatility across various industries. Featuring an intuitive interface with a clear LCD backlit display, installation and operation are a breeze, streamlining your monitoring process without the need for extensive training. Encased in a robust 1/2-DIN IP66 NEMA 4X enclosure, the 4000 SC analyzer is engineered for durability and flexibility, with advanced temperature compensation and multiple outputs to meet the most stringent monitoring requirements.

KEY FEATURES


  • Single-Channel Specific Conductivity for use with a wide range of contacting (specific) conductivity sensors

  • Ultra-Pure Water temperature compensation

  • Easy to install and configure with two relays, two digital inputs, and two 4-20mA outputs

  • Durable 1/2-DIN IP66 NEMA 4X enclosure can be panel, surface, or pipe-mounted

  • Simple, intuitive menu, with soft-touch buttons and a large, clear LCD backlit display

  • Multiple ranges from 0 to 9.999 µS/cm (Ultra-Pure Water) to 0 to 999.99 mS/cm (General Purpose)

  • Choice of 12-30VDC or 90-265VAC versions

MECHANICAL


Instrument housing

  • UL 94-V0 PC/ABS
  • Front panel 144 x 144mm (Panel-cut out: 138 x 138mm)
  • Depth behind panel: 77mm maximum 
  • Hinged / retained front

Ingress protection rating (IEC 60529 protection rating)

IP66

SAMPLE DELIVERY CONDITIONS


Ambient operating conditions 

Temperature -20 to +55°C, relative humidity 5 to 95%. non-condensing. 


Power supply

  • Universal 90-265vAC, 10W max.
  • LV Option 12-30vDC, 5W max.

PERFORMANCE


Display

3.75″, 240 x 128 dot, LCD module


Display backlight

It can be set to flash to indicate the instrument’s alarm status.


Languages

English, French, Spanish, and Italian


Buttons

5, silicone rubber tactile feedback micro-switched


Digital inputs

Two contact inputs for remote activation of user-defined operations. It can be configured to operate normally in open or closed modes. (All versions except the DOD17)


Current output user calibration 

± 0.01mA, 2 point 0/4 and 20 mA for remote monitor calibration.


Setpoint modes

High, low, band, delay timer adjustable from 00:00 to 59:59 mm:ss. Dose alarm timer, with supplementary initial charge function. Both adjustable from 00:00 to 59:59 mm:ss.


MicroSD card interface

Enables on site upgrading of instrument software. SD, SDHC and SDXC-FAT32 cards supported.


Measurement input

Any contacting conductivity cell. 


Connection cable

Up to 30 meters


Measurement ranges

  • 0-99.99 µS/cm to 0-999.9 mS/cm (K= 0.01 to 10.0)

Conductivity accuracy

± 0.5% of range.


Linearity and Repeatability

±0.1% of range.


Operator adjustment (conductivity)

± 10 % slope (gain) adjustment for solution calibration.


Range of temperature measurement

-50 °C to +150 °C (-58 °F to + 302 °F) for full specification.


Temperature accuracy

± 0.5 °C


Temperature sensor

Pt1000 RTD input. Up to 30 meters of cable. Temperature sensor can be mounted in the sensor or separately.