Waltron 9072 Hydrazine Analyzer

9072 Hydrazine Analyzer

Range: 0-1000ppb


The Waltron 9072 Hydrazine Analyzer delivers automatic, continuous measurement of hydrazine concentration independent of flow in clean water applications. Measurement is based on a highly-accurate potentiostatic electro-chemical principle. The 9072 is used for industrial process monitoring and control of residual hydrazine concentration in boiler feedwater, the primary loop of nuclear power plants, and water treatment plants. The potentiostatic technology allows for a much faster response time than amperometric or colorimetric systems, ensuring short start-up times, easy operation, and low maintenance costs.

  • Wide range analysis: 0-1000 μg/L (ppb)

  • Automatic temperature and flow compensation

  • Fast response time (t90 = 30 sec)

  • No zero point adjustment required

  • Withstands pressures up to 8 bar (116 psi)

  • Analog and digital interface; data logging option

  • Minimal maintenance required
PERFORMANCE

Range

0-1000 μg/L (ppb)


Accuracy

+/- 3% of reading


Response Time

+/- 3% of reading


Temperature Compensation

Automatic


Calibration

Process Calibration


Analog Outputs

0-20mA, 4-20mA (shunt max 500 Ohm)


Digital Outputs

Serial interface RS232


Alarms

Flow, calibration, power


Power

100VAC – 240 VAC, 50/60 Hz

SAMPLE DELIVERY OPERATING CONDITIONS

Sample Flow

3-18 L/h


Sample Temperature

32-140° F (0-60° C)


Ambient Temperature

32-131° F (0-55° C)


Pressure

≤ 8 bar (116 psig)


Sample Conductivity 

≥ 3 μS/cm; salt cell is required if conductivity is less

MECHANICAL

Dimensions

Height=25.75”(65.4 cm), Width=8”(20.3 cm), Depth=6.25”(15.9 cm)


Protection Class

IP54

Piping

6mm SS tube fittings

Materials used

Measuring Electrode: gold
Counter Electrode: stainless steel 1.4571 (314)
Reference Electrode: Ag/AgCl in saturated KCl-solution

Technical Documents