Waltron Waltron 9091C Dissolved Hydrogen Analyzer

9092 Free Chlorine Analyzer

Range: 0-1000µg/L(ppb)

Range: 0-10mg/L(ppm)


The Waltron 9092 Free Chlorine Analyzer delivers automatic, continuous measurement of free chlorine, active chlorine, ozone, chlorine dioxide and bromine in clean water applications. Measurement is based on a highly-accurate electro-chemical process. The 9092 is used for industrial process monitoring and process control of free chlorine in feed water, drinking water treatment, cooling water treatment, water management in swimming pools, or the residual concentration Cl2 of chlorine-treated clear waste water. The potentiostatic technology allows for a much faster response time than membrane or amperometric systems, ensuring short start-up times, easy operation, and low maintenance.

  • Wide range analysis: 0-1 mg/L (ppm), 0-10 mg/L (ppm)

  • Automatic temperature and flow compensation

  • Fast response time (t90 = 30 sec)

  • No zero point adjustment required

  • Withstands pressures up to 8 bar (116 psi)

  • Analog and digital interface; data logging option

  • Process calibration

  • Minimal maintenance required

  • Automatic pH compensation for free chlorine reading
PERFORMANCE

Range

0-1 mg/L (ppm), 0-10 mg/L (ppm)


Accuracy

+/- 3% of reading


Response Time

t90 = 30 sec


Temperature Compensation

Automatic


pH Compensation

Automatic


Calibration

Manual (Cl2 process calibration; pH calibration with 4 & 7 buffer solution)


Analog Outputs

0-20mA, 4-20mA (shunt max 500 Ohm)


Digital Outputs

Serial interface RS232


Alarms

Flow, calibration, power


Power

100VAC – 240 VAC, 50/60 Hz

SAMPLE DELIVERY OPERATING CONDITIONS

Sample Flow

3-18 L/h


Sample Temperature

32-140° F (0-60° C)


Ambient Temperature

32-131° F (0-55° C)


Pressure

0.3-8 bar (5-116 psig)


Sample Conductivity 

≥ 10 μS/cm; salt cell is required if conductivity is less

MECHANICAL

Dimensions

Height=25.75”(65.4 cm), Width=8”(20.3 cm), Depth=6.25”(15.9 cm)


Protection Class

IP54

Piping

6mm SS tube fittings

Materials used

Measuring Electrode: gold
Counter Electrode: stainless steel 1.4571 (314)
Reference Electrode: Ag/AgCl in saturated KCl-solution

Technical Documents