9095 Multi-Conductivity Analyzer
Range: 0-9.999µS/cm
Range: 0-99.9µS/cm
The Waltron 9095 Multi-Conductivity Analyzer continuously measures conductivity before and after cation exchange while automatically calculating the sample’s pH value and alkalizing the reagent concentration of feed water, steam, and condensate. The 9095 improves operational efficiency by providing comprehensive data and reducing the need for multiple instruments. An enhanced color touchscreen user interface makes the 9095 easy and intuitive to operate and configure. This combination measurement technology saves panel space and delivers the most accurate specific and cation conductivity measurements with simple maintenance and consumables.
- Highly accurate calculated pH
- Alarm for cation resin exhaustion
- Easy to replace cation column
- Easy to mount stainless steel panel
- One transmitter for two conductivity values and pH
- Dual channel option (9094C)
PERFORMANCE
Range
0-10µS/cm, 0-100µS/cm
Accuracy
< 0.6% of the measuring range
Reagent Consumption (approximate)
Cation Resin (pre-rinsed); replace cartridge every two months
Sensor Classification
Cationic exchange and conductivity sensors
Data Storage
Extract via USB 2.0 (optional)
Power Supply
AC 110 to 240 V +10 / -15%; 48 to 63 Hz or AC/DC 20 to 30 V; 48 t0 63 Hz
OPERATING CONDITIONS
Sample Temperature Range
32 – 122° F (0 – 50 °C)
Ambient Temperature Range
32 – 140° F (0 – 60 °C)
Inlet Sample Flowrate
.031 – .044 GPM (7-10L/h), 6mm Swagelok
Inlet Sample Pressure
Maximum of 145 psig (10 bar)
Outlet Sample
Atmospheric
SIGNAL OUTPUTS
Analog Outputs
5 x 0/4 – 20 mA for measured signals
Alarms
General Alarm, Resin Exhausted Alarm
Optional
RS422/RS485
MECHANICAL
Construction
Transmitter – high strength ABS, IP67
Wet section – stainless steel or PVC
Dimensions
14.5 x 33.5 x 10″ (368 x 851 x 254mm)
Mounting
Wall mount or Panel mount
Finish
Corrosion resistant