9165S Sequenced Luminescent Dissolved Oxygen Analyzer
Ranges: 0-20ppm, 0-2000 pp
The 9165S Sequenced Luminescent Dissolved Oxygen system, paired with our new sequencing system, utilizes the same reliable luminescent technology for measuring dissolved oxygen in water at ppb levels. Luminescent technology features superior benefits and cost savings compared to traditional dissolved oxygen sensing technologies. The analyzer system provides high accuracy and excellent long-term stability for up to four sample points.
KEY FEATURES
- Four sample analysis
- No sensor maintenance (no membrane, no electrolyte)
- No calibration or service for up to a year
- Measurement range 0-2000 ppb
- Extremely fast response time
- No sample interferences
- Excellent results independent of sample flow rate
- Simple operation
- Rugged design
- Four dedicated 4-20mA outputs
PERFORMANCE
Range
0-2000ppb; 0-45ppm
Accuracy
+/- 0.6ppb at 1ppb / 2% of the measured value
Sample Rate
Adjustable time on sample, sensor reading every 30 seconds
Units
PPB
Current Outputs
(4)4-20mA current output
Power
100-240 VAC
Calibration
Yearly sensor exchange program
Rinsedown
t90 = 3 min
SAMPLE DELIVERY OPERATING CONDITIONS
Sample Temperature
23 – 113°F (-5 – 40°C)
Ambient Temperature
32 – 131°F (0 – 55°C)
Sample Pressure
Maximum of 145 psig (10 bar)
Sample Interfaces
None
Sample Connection
1/4” OD SS Tubing (Swagelok)
Sample Medium
Liquid Water
MECHANICAL
Construction/Finish
Stainless Steel & Plastic / Corrosion Resistant
Materials in Contact w/ Sample
Stainless Steel
Dimensions
Sensor: Height = 33.5” (851 mm), Width = 12” (305 mm), Depth = 7.5” (191 mm)
Mounting
Wall mount
Degree of Protection
IP 67 Connections
Technical Documents