Waltron 4095 Multi-Conductivity Analyzer

4095 Multi-Conductivity Analyzer

Advanced Multi-Conductivity and pH Monitoring with Waltron’s 4095 Analyzer


The Waltron 4095 Multi-Conductivity Analyzer delivers innovative technology for precise water and steam monitoring tailored for power plants and industrial facilities. It offers unparalleled accuracy, streamlined single-transmitter displays, and easy data logging. Supported by Waltron’s strong U.S. team, the 4095 ensures reliable and efficient conductivity and pH analysis.


  • Conductivity measurement range: 0.055 to 1000 uS/cm
  • Calculation of pH value in the range from pH 7.5 to 11.5
  • Space-saving panel mount
  • Calculation of alkalizing reagent consumption, e.g. ammonia in the range from 0.01 to 10 ppm
  • Simultaneous measurement and display of specific and cation conductivity, pH, alkalizing reagent, sample temperature and sample flow
  • Temperature compensation preset for strong acids (wide range selectable for other sample conditions)
  • Optional 4010 resin cartridge assembly
  • Extremely low power consumption
  • 1/4-DIN cutout with color touchscreen
  • Connects to a wide variety of specific conductivity sensors

Specifications


  • Parameter: Specific/Cation/Calculated pH
  • Technology: Contacting
  • Range(s): 0-9.999 μS/cm, 0-99.99 μS/cm
  • Accuracy: (+/-) 1% of measured value or +/- 1 digit (whichever is greater)
  • Sample Streams: 1
  • Response Time: t90 < 30 sec
  • Sample Frequency: Continuous
  • Analog Outputs: 4 x 0/4 – 20 mA for measured signals
  • Digital Outputs: Optional RS422/RS485
  • Alarms: General Alarm
  • Calibration Method: Process
  • Ambient Conditions: 0-55°C / (32-131°F)
  • Sample Temperature: Probe-dependent
  • Sample Flow/Pressure: Probe-dependent

  • Power: 110 to 240 VAC, 48 to 63 Hz
  • IP Rating: IP 66, NEMA 4X
  • Mounting: Panel only
  • Materials: High Strength ABS

Coming Soon!

PERFORMANCE


Range

0 to 9 999 uS/cm, 0 to 99.99 uS/cm


Accuracy

< 0.6% of the measuring range


Degassing Efficiency

93.5% in less than 90 seconds


User Interface

5.5” color touch screen


Alarms

General Alarm, Low Degassing Efficiency Alarm, Resin Exhausted Alarm


Analog Outputs

4 x 0/4 – 20 mA for measured signals


Data Storage

Extract via USB 2.0


Power

110 to 240 VAC. 48 to 63 Hz


Options

Optional RS422/RS485


Method

Meets ASTM D4519-16 for degassing method


OPERATING CONDITIONS


Sample Temperature

32 – 140° F (0 – 60 °C)


Ambient Temperature

32 – 131° F (0 – 55 °C)


Sample Flow Rate

031 – .044 GPM (7-10L/h)


Pressure

Maximum of 145 psig (10 bar)


Sample Medium

Liquid


MECHANICAL


Construction

Transmitter – high strength ABS. IP67
Wet Section – stainless steel or PVC


Dimensions

29 x 37 x 10″ (613 x 939 x 254mm)


Mounting

Wall mount or panel mount


Finish

Corrosion resistant