9096 Degas Cation Conductivity Analyzer
Range: 0-9.999µS/cm
Range: 0-99.9µS/cm
The Waltron 9096 Degas Cation Conductivity Analyzer accurately measures specific, cation, and degassed cation conductivity (as well as calculated pH) in ultrapure water applications. The patented dynamic degassing module brings industry-leading degassing efficiency with actionable results in less than 90 seconds. An enhanced color touchscreen user interface makes the 9096 easy and intuitive to operate and configure, and ready-to-use consumables simplify maintenance tasks. This combination measurement technology brings the highest degassing efficiencies available on the market, delivers the most accurate measurements, and ensures 100% confidence in reducing turbine start-up times. This instrument uses ambient air forced against the flow of the water sample to achieve its high degassing efficiency, resulting in significantly faster indications than any other comparable systems available. The cost savings on each start-up cycle between the Waltron dynamic system and the fastest reboiler system are significant based on even the most conservative assumptions. Additional benefits are better energy efficiency and reduced emissions of heat and exhaust.
- Heatless dynamic degassing technology
- Industry-leading degassing efficiency
- Sample temperature fluctuation does not affect accuracy
- No gas emissions or volatile acids
- One transmitter for three conductivity values and pH
- Alarms for resin exhaustion and low degassing efficiency
- Easy to replace cation resin bottle
- Easy to mount stainless steel panel
- Significantly reduced turbine start-up times
PERFORMANCE
Range
0-10µS/cm, 0-100µS/cm
Accuracy
< 0.6% of the measuring range
Reagent Consumption (approximate)
Cation Resin (pre-rinsed) and Soda Lime; replace cartridges every two months
Sensor Classification
Cationic exchange and conductivity sensors
Degassing Efficiency
93.5%
Data Storage
Extract via USB 2.0
Power Supply
AC 110 to 240 V +10 / -15%; 48 to 63 Hz or AC/DC 20 to 30 V; 48 t0 63 Hz
OPERATING CONDITIONS
Sample Temperature Range
32 – 122° F (0 – 50 °C)
Ambient Temperature Range
32 – 140° F (0 – 60 °C)
Inlet Sample Flowrate
.031 – .044 GPM (7-10L/h), 6mm Swagelok
Inlet Sample Pressure
Maximum of 145 psig (10 bar)
Outlet Sample
Atmospheric
SIGNAL OUTPUTS
Analog Outputs
5 x 0/4 – 20 mA for measured signals
Alarms
General Alarm, Low Degassing Efficiency Alarm, Resin Exhausted Alarm
Optional
RS422/RS485
MECHANICAL
Construction
Transmitter – high strength ABS, IP67
Wet section – stainless steel or PVC
Dimensions
29 x 37 x 10″ (613 x 939 x 254mm)
Mounting
Wall mount or Panel mount
Finish
Corrosion resistant