9135 Sodium Analyzer

Range: 0.1ppb – 10ppm


Waltron’s 9135 Sodium Analyzer delivers exceptional results in a streamlined package. An enhanced color touch screen user interface makes the 9135 easy to operate and easy to maintain. The 9135 Sodium Analyzer can be used in a variety of ultrapure water applications. Common measuring points in a boiler system for online sodium measurement include: make-up water, condensate, boiler feedwater and saturated steam/main steam. The Waltron 9135 Sodium Analyzer comes with true fully automatic calibration and offers many benefits in terms of ease-of-use and increased functionality.

  • Larger user-friendly color touch screen interface

  • Wide analysis range: 0.1ppb – 10ppm

  • Calibration: automatic 2-point or single-point; user may select weekly, bi-monthly, or monthly calibration frequency as well as day and time

  • Hassle-free maintenance: calibration standards and electrode refill can last up to 3 months

  • Low sample flow indication (optional)

  • Quick connect electrodes and probe positioning clamps

  • Automatic reference electrode electrolyte refill system

  • Grab sample capability

  • Two 4-20 mA isolated current outputs

  • Internal Data Logger set to store readings every 10 minutes for 30 days

  • Automatically logs last 10 alarms and results of last 10 calibrations

  • Reagents: Ammonium Hydroxide (standard) or Di-isopropylamine (optional)

Range

0.1 ppb – 10 ppm


Accuracy

± 5% of reading or ± 0.1 ppb (whichever is greater) within ± 5°C of calibration temperature


Reproducibility 

± 5% of reading or ± 0.1 ppb (whichever is greater) at constant temperature


Response Time (approximate)

90% of 10-1 ppb step: 3 minutes; 90% of 100-1 ppb step: 5 minutes


Temperature Compensation

Automatic


pH Compensation

Alkaline vapor buffer: 0.88 SG (28% to 30% NH3) ammonia above 0.1 ppb sodium, 99% di-isopropylamine solution below 0.1 ppb sodium


Digital Output

Communication interface via RS-232 & RS-485 using MODBUS RTU protocol


Alarms

Low flow alarm and three additional voltage-free contacts, alarm points set from transmitter rated at 2A, 250 VAC (non-inductive)


Power

Wide range of input power supply 90 VAC – 250 VAC


Calibration

Automatic 2-point or single-point; user may select weekly, bi-monthly, or monthly calibration frequency as well as day and time


Certification (on request)

Meets low voltage and low electromagnetic compatibility directives


Sample 

Temperature: 41 – 131°F (5 – 55°C); Flow: 150 – 400 ml/min


Ambient Temperature

32 – 131°F (0 – 55°C)


Pressure

5 – 30 psig (0.35 – 2.07 bar)


Composition

Sample should be filtered to 60 microns, free of film forming compounds


Construction

Display – high strength ABS, IP65 (NEMA 4x)

Enclosure – stainless steel sheet metal, all wetted parts stainless steel or PVC


Dimensions

Height=26”(66.1 cm), Width=15”(38.1 cm), Depth=12”(30.5 cm)


Weight

33 lbs (15 kg)


Wiring

Eight (8) conduit knockouts. Quick connect electrodes.


Piping

Sample Inlet: 1/4″ (0.635 cm) OD Swagelok; Drain: 3/8″ (0.953 cm) OD tube


Distance

Up to 325 ft (100 meters) on request


Finish

Corrosion resistant

Technical Documents