9165S LDO Smart Sensor Analyzer

9165 Dissolved Oxygen Analyzer

Range: 0-45ppm

Range: 0-2000ppb


The Waltron 9165 Luminescent Dissolved Oxygen Analyzer (LDO) utilizes optical (luminescent) technology for measuring dissolved oxygen in water at extremely low to sub-ppb levels, independent of flow. Luminescent technology features superior benefits and cost savings compared to traditional dissolved oxygen sensing technologies, including ease of maintenance and operation. The 9165 provides high accuracy with excellent long-term stability and can be used in a variety of online analysis applications throughout many different industries.

KEY FEATURES


  • No sensor maintenance (no membrane, no electrolyte)

  • Wide measurement range: 0–2000ppb; 0-45ppm

  • No calibration for up to 2 years

  • Extremely fast response time with minimal flow requirements

  • No sample interferences

  • Independent configuration of ranges, alarms, and reading outputs

  • One month of internal data logging, USB download

  • Simple system setup

  • Rugged and compact design

  • Plug-and-play addition of multiple sensors

  • Up to four sample streams on one electronics platform

  • Complete touchscreen control with sample point labeling

  • Dynamic sampling rates

PERFORMANCE


Range

0-2000ppb; 0-45ppm


Accuracy

± 0.5 ppb + 2% of the measured value


Resolution

0.1 ppb


Units

ppb, ppm, % O2, % a.s. ppm vol/vol


Sample Rate

One sample every 10 – 999 sec (adjustable)


Current Outputs

Four configurable 4-20mA current (analog) outputs


Digital Output

Communication interface using RS-232


Alarms

(4) NC or NO, General Alarm, Temperature Alarm, High or Low Concentration Alarms


Power

100 – 250 VAC ~ 50 – 60 Hz, optional 24V DC, 25W


Options

CE Certification; Profibus; multiple sensor configuration; certificate of calibration


Options

Profibus; multiple sensor configuration; certificate of measurement

SAMPLE DELIVERY OPERATING CONDITIONS


Sample Temperature

-5 – 70°C

Steam Sterilization up to 130°C


Ambient Temperature

32 – 131°F (0 – 55°C)


Pressure

Maximum of 145 psig (10 bar)


Sample Connection

1/4” (6 mm) Swagelok


Sample Medium

Liquid


Sample Memory

Up to 400 measurements

MECHANICAL


Construction

Transmitter – high strength ABS, IP65 (NEMA 4x)

Wet section – high strength ABS and sheet metal, all wetted parts stainless steel or PVC, IP55 (NEMA 4)


Dimensions

Combined Analyzer: Height=25.75”(65.4 cm), Width=8”(20.3 cm), Depth=6.25”(15.9 cm)

Separated Transmitter: Height=10.92”(27.74 cm), Width=7” (17.78 cm), Depth=7.63”(8.99 cm)

Separated Wet Section: Height = 10.5” (26.7 cm), Width = 4.5” (11.4 cm), Depth = 3.5” (8.9 cm)


Mounting

Wall mount or Panel mount


Wiring

IP 67 Connections


Distance

Up to 49ft (15 meters) on request


Finish

Corrosion resistant

Technical Documents